E Series Multifunction DAQ
16-Bit, 333 kS/s, 16 Analog Inputs
201
National Instruments • Tel: (800) 433-3488 Fax: (512) 683-9300 [email protected] ni.com
E Series 16-Bit Multifunction DAQ
DAQ and Signal Conditioning
automotive testing and laboratory measurements. The
NI DAQPad-6052E can be powered by any 9 to 25 VDC power
supply such as a car battery. The unit comes with a universal AC-
to-DC power supply that can interface to standard power outlets
anywhere in the world. You can also use the optional BP-1
rechargeable battery pack to power the device for up to 2.5 hours.
Refer to page 215 for a more comprehensive look at our portable DAQ
products as well as portable DAQ accessories.
Driver Software
NI-DAQ is the robust driver software included with all National
Instruments data acquisition and signal conditioning products.
This easy-to-use software tightly integrates the full functionality of
your DAQ hardware to LabVIEW, LabWindows/CVI, and
Measurement Studio for Visual Basic. High-performance features
include multidevice synchronization, networked measurements,
and DMA data management. Bundled with NI-DAQ, the
Measurement & Automation Explorer utility simplifies the
configuration of your measurement hardware with device test
panels, interactive measurements, and scaled I/O channels.
NI-DAQ also provides numerous example programs for LabVIEW
and other application development environments to get you
started with your application quickly.
Services and Support/Training
As a complement to your data acquisition and signal conditioning
product, consider:
•Technical Support – Included in hardware/software purchase
through Applications Engineers worldwide, Web resources
with more than 1000 example programs and more than 7000
KnowledgeBases, and Premier Support – ni.com/support
NI Factory Installation Services (FIS) – Software and hardware
installed in PXI and PXI/SCXI systems, tested and ready to use
ni.com/advisor
Calibration – Includes NIST-traceable basic calibration
certificate, services for ANSI/NCSL-Z540 and periodic
calibration – ni.com/calibration
Extended Warranty – Meet project life-cycle requirements and
maintain optimal performance in a cost-effective way –
ni.com/services
Data Acquisition Training – Instructor-led courses – ni.com/training
•Professional Services – Feasibility, consulting, and integration
through our Alliance Program members – ni.com/alliance
For more information on NI services and support, visit ni.com/services
Table 2. NI 6052E Analog Input Accuracy Specifications
Absolute Accuracy Relative Accuracy
Nominal Range (V) % of Reading Offset Noise + Quantization (µV) Temp Absolute Accuracy Resolution (µV)
Positive FS Negative FS 24 Hrs 1 Year (µV) Single Pt. Averaged Drift (%/°C) at Full Scale (mV) Single Pt. Averaged
10.0 -10.0 0.0354 0.0371 947.0 981.0 87.0 0.0006 4.747 1145.0 115.0
5.0 -5.0 0.0054 0.0071 476.0 491.0 43.5 0.0001 0.876 573.0 57.3
2.5 -2.5 0.0354 0.0371 241.0 245.0 21.7 0.0006 1.190 286.0 28.6
1.0 -1.0 0.0354 0.0371 99.2 98.1 8.7 0.0006 0.479 115.0 11.5
0.5 -0.5 0.0354 0.0371 52.1 56.2 5.0 0.0006 0.243 66.3 6.6
0.25 -0.25 0.0404 0.0421 28.6 32.8 3.0 0.0006 0.137 39.2 3.9
0.1 -0.1 0.0454 0.0471 14.4 22.4 2.1 0.0006 0.064 27.7 2.8
0.05 -0.05 0.0454 0.0471 9.7 19.9 1.9 0.0006 0.035 25.3 2.5
10.0 0.0 0.0054 0.0071 476.0 491.0 43.5 0.0001 1.232 573.0 57.3
5.0 0.0 0.0354 0.0371 241.0 245.0 21.7 0.0006 2.119 286.0 28.6
2.0 0.0 0.0354 0.0371 99.2 98.1 8.7 0.0006 0.850 115.0 11.5
1.0 0.0 0.0354 0.0371 52.1 56.2 5.0 0.0006 0.428 66.3 6.6
0.5 0.0 0.0404 0.0421 28.6 39.8 3.0 0.0006 0.242 48.2 3.9
0.2 0.0 0.0454 0.0471 14.4 22.4 2.1 0.0006 0.111 27.7 2.8
0.1 0.0 0.0454 0.0471 9.7 19.9 1.9 0.0006 0.059 25.3 2.5
Note: Accuracies are valid for measurements following an internal E Series calibration. Averaged numbers assume dithering and averaging of 100 single-channel readings. Measurement accuracies are listed for
operational temperatures within ±1 °C of internal calibration temperature and ±10 °C of external or factory-calibration temperature. One-year calibration interval recommended. The absolute accuracy at full scale
calculations were performed for a maximum range input voltage (for example, 10 V for the ±10 V range) after one year, assuming 100 pt averaging of data. See overview on page 194 for example calculations.
Absolute Accuracy
Nominal Range (V) % of Reading Temp Absolute Accuracy
Positive FS Negative FS 24 Hrs 90 Days 1 Year
Offset
(µV) Drift (%/°C) at Full Scale (mV)
10 -10 0.0044 0.0052 0.0061 798 0.0001 1.405
10 0 0.0044 0.0052 0.0061 569 0.0001 1.176
Note: Temp Drift applies only if ambient is greater than ±10 °C of previous external calibration. See page 194 for example calculations.